Notice Information
Notice Title
Quantum Cascade Laser Analyser for Greenhouse Gases and Air Pollutants
Notice Description
A rack mounted quantum cascade laser instrument capable of high precision (< 0.5 nmol mol-1 for all except carbon dioxide <= 25 nmol mol-1) simultaneous measurements of carbon dioxide, methane nitrous oxide, carbon monoxide, ammonia, nitrogen monoxide, nitrogen dioxide and sulfur dioxide in the 0 - 10 umol mol-1 range (0 - 30 umol mol-1 for methane and 0- 5000 umol mol-1 for carbon dioxide).
Lot Information
Lot 1
Multicomponent quantum cascade laser analyser with >= 1 Hz data acquisition rate that allows the simultaneous measurement of carbon dioxide (CO2), methane (CH4), nitrous oxide (N2O), carbon monoxide (CO), ammonia (NH3), nitrogen monoxide (NO), nitrogen dioxide (NO2), sulphur dioxide (SO2) with the following amount fractions ranges and one sigma (1s) precisions: Carbon dioxide (amount fraction range: 0 - 5000 umol mol-1; precision 25 nmol mol-1), methane (amount fraction range: 0 - 30 umol mol-1; precision 0.22 nmol mol-1), nitrous oxide (amount fraction range: 0 - 10 umol mol-1; precision 0.1 nmol mol-1), carbon monoxide (amount fraction range: 0 - 10 umol mol-1; precision 0.01 nmol mol-1), ammonia (amount fraction range: 0 - 10 umol mol-1; precision 0.5 nmol mol-1), nitrogen monoxide (amount fraction range: 0 - 10 umol mol-1; precision 1 % or 1 nmol mol-1, whichever is greater), nitrogen dioxide (amount fraction range: 0 - 10 umol mol-1; precision 1 % or 1 nmol mol-1, whichever is greater) and sulphur dioxide (amount fraction range: 0 - 10 umol mol-1; precision 1 % or 1 nmol mol-1, whichever is greater). The instrument shall have a flow rate of 0.5 - 1.5 slpm, a leak rate of <= 10-9 mbar xl/s and response time of < 2 minutes. It shall be able to operate under environmental conditions of ambient temperature: 150C to 300C and ambient humidity: 30-80% RH without condensing. The wetted materials shall be corrosion resistance preferable SS 316/316L. A gas connection of 1/4" or 6mm Swagelok is required with an inlet pressure of 1 - 5 bar absolute. The instrument shall have the following connectivity: 100-1000 Base-T Ethernet, USB 2.0 or better, serial RS232/485, Modbus TCP and an isolated 4-20 mA or 0-10 Vdc analogue output. Power consumption ideally should < 500 W and with a total weight of < 50 kg. The analyser shall be CE/UKCA certificated with a minimum of 12-month warranty.
Procurement Information
Miro Analytical AG is the only vendor that can provide an analytical solution that enables the simultaneous measurement of all the major greenhouse gases (CO2, CH4, N2O and CO) and air quality pollutants (NO, NO2, SO2 and NH3) with the required high precision and broad amount fraction ranges that are required. Unique features include: 1. The ability to measure all these compounds simultaneously due to a unique dual colour laser system. 2. The achieved precisions with this analyser match or exceed other optical spectrometers that are typically measuring one or two compounds. 3. The range of amount fractions that are possible to measure exceed those typical of other optical spectrometers that are measuring one or two compounds.
Notice Details
Publication & Lifecycle
- Open Contracting ID
- ocds-h6vhtk-03eb22
- Publication Source
- Find A Tender Service
- Latest Notice
- https://www.find-tender.service.gov.uk/Notice/021940-2023
- Current Stage
- Award
- All Stages
- Award
Procurement Classification
- Notice Type
- Award Notice
- Procurement Type
- Standard
- Procurement Category
- Goods
- Procurement Method
- Limited
- Procurement Method Details
- Award procedure without prior publication of a call for competition
- Tender Suitability
- Not specified
- Awardee Scale
- Large
Common Procurement Vocabulary (CPV)
- CPV Divisions
38 - Laboratory, optical and precision equipments (excl. glasses)
-
- CPV Codes
38000000 - Laboratory, optical and precision equipments (excl. glasses)
Notice Value(s)
- Tender Value
- Not specified
- Lots Value
- Not specified
- Awards Value
- Not specified
- Contracts Value
- £205,274 £100K-£500K
Notice Dates
- Publication Date
- 28 Jul 20232 years ago
- Submission Deadline
- Not specified
- Future Notice Date
- Not specified
- Award Date
- 27 Jul 20232 years ago
- Contract Period
- Not specified - Not specified
- Recurrence
- Not specified
Notice Status
- Tender Status
- Complete
- Lots Status
- Not Specified
- Awards Status
- Active
- Contracts Status
- Active
Buyer & Supplier
Contracting Authority (Buyer)
- Main Buyer
- NATIONAL PHYSICAL LABORATORY
- Contact Name
- Not specified
- Contact Email
- charley.choules@npl.co.uk
- Contact Phone
- Not specified
Buyer Location
- Locality
- TEDDINGTON
- Postcode
- TW11 0LW
- Post Town
- Twickenham
- Country
- England
-
- Major Region (ITL 1)
- TLI London
- Basic Region (ITL 2)
- TLI7 Outer London - West and North West
- Small Region (ITL 3)
- TLI75 Hounslow and Richmond upon Thames
- Delivery Location
- Not specified
-
- Local Authority
- Richmond upon Thames
- Electoral Ward
- Teddington
- Westminster Constituency
- Twickenham
Further Information
Notice URLs
Open Contracting Data Standard (OCDS)
View full OCDS Record for this contracting process
The Open Contracting Data Standard (OCDS) is a framework designed to increase transparency and access to public procurement data in the public sector. It is widely used by governments and organisations worldwide to report on procurement processes and contracts.
{
"tag": [
"compiled"
],
"id": "ocds-h6vhtk-03eb22-2023-07-28T10:25:19+01:00",
"date": "2023-07-28T10:25:19+01:00",
"ocid": "ocds-h6vhtk-03eb22",
"initiationType": "tender",
"tender": {
"id": "ocds-h6vhtk-03eb22",
"legalBasis": {
"id": "32014L0024",
"scheme": "CELEX"
},
"title": "Quantum Cascade Laser Analyser for Greenhouse Gases and Air Pollutants",
"status": "complete",
"classification": {
"scheme": "CPV",
"id": "38000000",
"description": "Laboratory, optical and precision equipments (excl. glasses)"
},
"mainProcurementCategory": "goods",
"description": "A rack mounted quantum cascade laser instrument capable of high precision (< 0.5 nmol mol-1 for all except carbon dioxide <= 25 nmol mol-1) simultaneous measurements of carbon dioxide, methane nitrous oxide, carbon monoxide, ammonia, nitrogen monoxide, nitrogen dioxide and sulfur dioxide in the 0 - 10 umol mol-1 range (0 - 30 umol mol-1 for methane and 0- 5000 umol mol-1 for carbon dioxide).",
"lots": [
{
"id": "1",
"description": "Multicomponent quantum cascade laser analyser with >= 1 Hz data acquisition rate that allows the simultaneous measurement of carbon dioxide (CO2), methane (CH4), nitrous oxide (N2O), carbon monoxide (CO), ammonia (NH3), nitrogen monoxide (NO), nitrogen dioxide (NO2), sulphur dioxide (SO2) with the following amount fractions ranges and one sigma (1s) precisions: Carbon dioxide (amount fraction range: 0 - 5000 umol mol-1; precision 25 nmol mol-1), methane (amount fraction range: 0 - 30 umol mol-1; precision 0.22 nmol mol-1), nitrous oxide (amount fraction range: 0 - 10 umol mol-1; precision 0.1 nmol mol-1), carbon monoxide (amount fraction range: 0 - 10 umol mol-1; precision 0.01 nmol mol-1), ammonia (amount fraction range: 0 - 10 umol mol-1; precision 0.5 nmol mol-1), nitrogen monoxide (amount fraction range: 0 - 10 umol mol-1; precision 1 % or 1 nmol mol-1, whichever is greater), nitrogen dioxide (amount fraction range: 0 - 10 umol mol-1; precision 1 % or 1 nmol mol-1, whichever is greater) and sulphur dioxide (amount fraction range: 0 - 10 umol mol-1; precision 1 % or 1 nmol mol-1, whichever is greater). The instrument shall have a flow rate of 0.5 - 1.5 slpm, a leak rate of <= 10-9 mbar xl/s and response time of < 2 minutes. It shall be able to operate under environmental conditions of ambient temperature: 150C to 300C and ambient humidity: 30-80% RH without condensing. The wetted materials shall be corrosion resistance preferable SS 316/316L. A gas connection of 1/4\" or 6mm Swagelok is required with an inlet pressure of 1 - 5 bar absolute. The instrument shall have the following connectivity: 100-1000 Base-T Ethernet, USB 2.0 or better, serial RS232/485, Modbus TCP and an isolated 4-20 mA or 0-10 Vdc analogue output. Power consumption ideally should < 500 W and with a total weight of < 50 kg. The analyser shall be CE/UKCA certificated with a minimum of 12-month warranty.",
"hasOptions": false
}
],
"items": [
{
"id": "1",
"deliveryAddresses": [
{
"region": "UK"
}
],
"relatedLot": "1"
}
],
"procurementMethod": "limited",
"procurementMethodDetails": "Award procedure without prior publication of a call for competition",
"procurementMethodRationaleClassifications": [
{
"scheme": "TED_PT_AWARD_CONTRACT_WITHOUT_CALL",
"id": "D_OUTSIDE_SCOPE",
"description": "The procurement falls outside the scope of application of the directive"
}
],
"procurementMethodRationale": "Miro Analytical AG is the only vendor that can provide an analytical solution that enables the simultaneous measurement of all the major greenhouse gases (CO2, CH4, N2O and CO) and air quality pollutants (NO, NO2, SO2 and NH3) with the required high precision and broad amount fraction ranges that are required. Unique features include: 1. The ability to measure all these compounds simultaneously due to a unique dual colour laser system. 2. The achieved precisions with this analyser match or exceed other optical spectrometers that are typically measuring one or two compounds. 3. The range of amount fractions that are possible to measure exceed those typical of other optical spectrometers that are measuring one or two compounds."
},
"awards": [
{
"id": "021940-2023-1",
"relatedLots": [
"1"
],
"title": "Quantum Cascade Laser Analyser for Greenhouse Gases and Air Pollutants",
"status": "active",
"hasSubcontracting": true,
"suppliers": [
{
"id": "GB-FTS-88574",
"name": "Miro Analytical AG"
}
]
}
],
"parties": [
{
"id": "GB-FTS-155",
"name": "National Physical Laboratory",
"identifier": {
"legalName": "National Physical Laboratory"
},
"address": {
"streetAddress": "Hampton Road",
"locality": "Teddington",
"region": "UK",
"postalCode": "TW11 0LW",
"countryName": "United Kingdom"
},
"contactPoint": {
"email": "charley.choules@npl.co.uk"
},
"roles": [
"buyer"
],
"details": {
"url": "http://www.npl.co.uk",
"classifications": [
{
"scheme": "TED_CA_TYPE",
"id": "BODY_PUBLIC",
"description": "Body governed by public law"
},
{
"scheme": "COFOG",
"description": "Research"
}
]
}
},
{
"id": "GB-FTS-88574",
"name": "Miro Analytical AG",
"identifier": {
"legalName": "Miro Analytical AG"
},
"address": {
"streetAddress": "Widenholzstrasse 1, CH-8304 Wallisellen, Switzerland",
"locality": "Wallisellen",
"region": "CH",
"postalCode": "CH-8304",
"countryName": "Switzerland"
},
"roles": [
"supplier"
],
"details": {
"scale": "large"
}
},
{
"id": "GB-FTS-156",
"name": "NPL Management Ltd",
"identifier": {
"legalName": "NPL Management Ltd"
},
"address": {
"streetAddress": "Hampton Road",
"locality": "Teddington",
"postalCode": "TW11 0LW",
"countryName": "United Kingdom"
},
"roles": [
"reviewBody"
]
}
],
"buyer": {
"id": "GB-FTS-155",
"name": "National Physical Laboratory"
},
"contracts": [
{
"id": "021940-2023-1",
"awardID": "021940-2023-1",
"title": "Quantum Cascade Laser Analyser for Greenhouse Gases and Air Pollutants",
"status": "active",
"value": {
"amount": 205274,
"currency": "CHF"
},
"dateSigned": "2023-07-28T00:00:00+01:00"
}
],
"language": "en"
}